 An
Energy Saving
Support Tool
3331, 3332 POWER HiTESTER
Power measuring instruments
Accurate evaluation of consumption power of electrical products
As efficient use of energy for household and office equipment becomes more and more essential, the new 3332 POWER HiTESTER does the job by offering a wide range of power measurement from standby to normal usage. The 3331 POWER HiTESTER is capable of evaluating 3-phase devices, such as industrial air conditioners and refrigerators, or single-phase, large-scale devices. Both power testers deliver high accuracy of 0.2% (45 to 66Hz), direct input up to 50A, and a broad bandwidth from 1Hz (the 3331 from 10Hz) to 100kHz. System construction is made easy with these compact, lightweight and reasonably priced tools, which come equipped with an external interface as a standard feature. The 3331 and 3332 can be used as a measuring component for a wide range of purposes, from research and development to equipment evaluation.
3332: Single-phase, 2-wire type that can accurately measure even standby power 3331: Single-phase, 3-wire and three-phase, 3-wire type for measuring power of large-scale equipment
1
Measurement from minute single-phase power to large-scale 60 kW 3-phase equipment.
A single-phase power meter compatible with devices with intermittent oscillation in broadband starting from 1Hz. from standby to usage
The 3332 covers a wide range of power measurement
Model Measurement line U Range I Range
Frequency characteristics
3332
1o2W only
3331
1o2W to 3o3W
15 to 600V (6 ranges) 150 to 600V (3 ranges) 1m to 50A (15 ranges) 500m to 50A (7ranges) 1Hz to 100kHz 10Hz to 100kHz 0.1%rdg.0.1%f.s. (45 to 66Hz) Approx. 210W 100H 261D mm
Basic accuracy Dimensions
The 3331 is compatible from single-phase to 3-phase devices
Range Table
The values in the shaded areas show the common range for the 3331 and 3332. Values in the ( mode and the single-phase, 3-wire mode (SUM display) of the 3331. ) show the range of the 3-phase, 3-wire
Voltage/Current can be viewed for each phase of the 3-phase line
U
I
1.0000mA 2.0000mA 5.0000mA 10.000mA 20.000mA 50.000mA 100.00mA 200.00mA 15.000 mW 30.000 mW 75.000 mW 150.00 mW 300.00 mW 750.00 mW 30.000 mW 60.000 mW 150.00 mW 300.00 mW 600.00 mW 60.000 mW 120.00 mW 300.00 mW 600.00 mW 150.00 mW 300.00 mW 750.00 mW 300.00 mW 600.00 mW 600.00 mW 500.00mA 7.5000 W 15.000 W 30.000 W 75.000 (150.00) W 1.2000 W 1.0000A 15.000 W 30.000 W 60.000 W 1.5000 W 3.0000 W 1.5000 W 3.0000 W 6.0000 W 1.2000 W 3.0000 W 6.0000 W 12.000 W 1.5000 W 3.0000 W 7.5000 W 15.000 W 30.000 W 1.5000 W 3.0000 W 6.0000 W 15.000 W 30.000 W 60.000 W 20.000A 300.00 W 600.00 W 1.2000kW 3.0000 W 6.0000 W 12.000 W 30.000 W 60.000 W 120.00 W 50.000A 750.00 W 1.5000kW 3.0000kW
15.000V 30.000V 60.000V 150.00V 300.00V 600.00V
U
I
2.0000A 30.000 W 60.000 W 120.00 W
5.0000A 75.000 W 150.00 W 300.00 W
10.000A 150.00 W 300.00 W 600.00 W
15.000V 30.000V 60.000V 150.00V 300.00V 600.00V
150.00 (300.00) W 300.00 (600.00) W 750.00 (1.5000 k) W 1.5000 (3.0000) kW 3.0000 (6.0000) kW 7.5000 (15.000) kW
150.00 (300.00) W 300.00 (600.00) W 600.00 (1.2000 k) W 1.5000 (3.0000) kW 3.0000 (6.0000) kW 6.0000 (12.000) kW 15.000 (30.000) kW 300.00 (600.00) W 600.00 (1.2000 k) W 1.2000 (2.4000) kW 3.0000 (6.0000) kW 6.0000 (12.000) kW 12.000 (24.000) kW 30.000 (60.000) kW
For apparent power and reactive power, the unit of watts in the above table is replaced by VA and var respectively.
2
Basic Performance of the 3331/3332
Evaluation of electric equipment such as inverters
High basic accuracy of 0.2%
More precise measurement non possible with a basic accuracy of 0.1% rdg.0.1% f.s.
Broadband feature compatible with frequency control devices
Wide range from 1Hz (the 3331 from 10Hz) to 100kHz is included for supporting measurement of inverters.
Responsitivity that follows transient power fluctuations
A achieve responses under 0.2 seconds for measurements of transient power fluctuations (Response speed set at FAST).
50A direct input
Measurement of large capacity equipment possible.
Simultaneous integration of current and power at a 6-digit high-resolution state
A maximum of 999999 (MWh or MAh) or up to a maximum of 10000 hours (416 days) of integration.
Measuring the effective value of basic wave components only
The average rectified effective value indicator method with a 500Hz low-pass filter can be selected.
Current waveform peak measurement function
The current waveform wave peak value and the maximum effective value can be detected.
Systems can be easily constructed
A compact design that fits a half-rack (rack-mount models also available at special order) GP-IB / RS-232C: Data can be transferred to a printer or computer for efficient data management. EXT.I/O (External input/output terminal): External control of integration START/STOP, and analog/monitor/D/A output can be performed for voltage/current/power parameters.
3332
3331
Accuracy
Frequency
(at 23C5C (73F9F), 80% rh sine wave input, power factor = 1, after 30 minutes warming-up time.) 3331: 23C3C (73F5.4F) (Period for which accuracy is guaranteed: 6 months)
Voltage*1 / Current / Active power
Input current 20 A or less 20A to 30A
1Hz to 2Hz*2*3 2Hz to 5Hz*2*3 5Hz to 10Hz*2*3 10Hz to 20Hz*2*4 20Hz to 30Hz*2*5 30Hz to 45Hz*2 45Hz to 66Hz 66Hz to 500Hz 500Hz to 1kHz 1kHz to 4kHz 4kHz to 8kHz 8kHz to 10kHz*6 10kHz to 20kHz 20kHz to 50kHz
50kHz to 100kHz*7
12%f.s. 5%f.s. 1.5%f.s. 1.0%f.s. 0.5%f.s. 0.1%rdg.0.2%f.s. 0.1%rdg.0.1%f.s. 0.1%rdg.0.2%f.s. 0.3%rdg.0.2%f.s. 0.3%rdg.0.2%f.s. 1.0%f.s. 1.0%f.s. 2.0%f.s. 5.0%f.s. 15.0%f.s.
2.0%f.s. 2.0%f.s. 2.0%f.s.
undefined undefined undefined
30A to 50A undefined undefined undefined undefined undefined undefined
*1 *2 *3 *4 *5 *6 *7
Voltage accuracy is the same as when current input is less than 20A. Measurement accuracy when response time is set to SLOW. Accuracy guaranteed for the 3332 only. Voltage/current accuracy for the 3331 is 2.0% f.s. Voltage/current accuracy for the 3331 is 1.0% f.s. Guaranteed accuracy for the 3331 is 2.0% f.s. Current is defined for 10A and less.
0.2%f.s.
undefined undefined undefined undefined undefined undefined undefined undefined
Basis of calculation
ch 1 2 SUM
Calculations of the 3331 3-phase, 3-wire mode are as follows in the table below. However, only SUM is displayed. Calculations for the 3332 follow the values for ch 1 in the table below. U, I, and P respectively indicate measured values of voltage, Active power (P ) Apparent power (S ) Reactive power (Q ) Power factor () Phase angle (o) current, and active power. However, values are not rounded for display (error: 1 dgt.). s indicates phase polarity, and is -1 when the current phase leads voltage, and +1 when it lags voltage.
3
From minute standby power to rush current for motors
Applications that efficiently evaluate electrical equipment
Common features of the 3331 and 3332
Measurement of rush current during device start-up
Measurement of the current waveform wave peak is possible, and if the peak hold function is used, wave peak detection of the motor rush current waveform (Max. 90A) and the maximum value of the effective value can be done.
Measurement of the RMS value for industrial frequency components.
The average rectified effective value indicator method (MEAN) with 500Hz low-pass filter is employed to measure basic wave RMS values of PWM voltage form inverters.
Comparison of voltage display values of RMS and MEAN with the low-pass filter
RMS display of PWM voltage waveforms. MEAN display of voltage waveform after filtering.
RMS response at 0.2 sec.
Understanding consumptive and regenerative conditions
Consumptive (+), regenerative (-), and total power integration values can be simultaneously measured on equipment that regenerate power.
Integration progress time
Total () integration value
Consumptive element (+) integration value
Regenerative element (-) integration value
Special Features of the 3331
3-phase line imbalance can be checked
The third phase voltage and current that had not been measured for 2-power measurement systems (2 voltage/2 current) can be calculated by vector calculation and displayed.
U1 voltage display U2 voltage display
The 3331 can be used to measure single phase power for 2 devices
Two devices with single-phase, 2-wire loads can be simultaneously measured, contributing to lower facilities cost.
Note) Ranges for each channel cannot be independently set.
U3 voltage display
Average voltage display for each phase
Special Features of the 3332
Measurement of standby power under 1W
Current input 1W or less can be precisely measured (guaranteed accuracy range is from 7.5mW) by employing the CT method (input resistance under 2m) for minimal instrument damage for current input, and with 150.000mW (150V-1mA range) in full scale as the highest sensitivity range for 100V devices. In addition, with guaranteed accuracy from 1Hz, the 3332 is also compatible with intermittent oscillating devices, such as videos.
Precise calculations of minute standby power
When the 150V/1.0000mA range is selected, integration from 000.000mWh can be performed. Low numbers/units are automatically switched to a 6-digit display, allowing measurement in high resolution.
Integration display screen
Current integration display
Integration progress time
Comparative decision function that can be used on the production line
Two items can be chosen from among voltage, current, power (active, reactive, apparent), peak current, power factor, phase angle, frequency and integration value for simultaneous comparative decisions. In addition to Hi/In/Lo LED lamps, decision results are output to contact points. Up to 10 conditions can be stored, a powerful function to reduce repetitive steps on small output/multi-product lines.
Other parameters
Active power integration
Comparator setting screen
Comparative value setting
Other parameters
Comparator decisions
Measurement value
4
Application example using the GP-IB/RS-232C interface
The 3331 and 3332 are equipped with the GP-IB and RS-232C interfaces as standard features, allowing complete control from a computer (except for turning the power supply ON/OFF). In addition, measurement data can be directly downloaded into commercially available spreadsheet software on a computer using application software, making the troublesome creation of test result charts easy, and supporting effective data management.
Example of chart calculation operation with EXCEL*
*EXCEL is a registered trademark of Microsoft Corporation.
Keeping standby power of household electrical equipment in the range of 1W and below
A plan for reducing contributions to global warming by raising the efficiency of electrical energy used by household and office equipment must take standby power into consideration. In the Japanese domestic market, precise measuring of minute standby power is needed, especially for manufacturers of audio/video devices, in order to follow a policy of keeping standby power under 1W.
Raising the efficiency of electrical energy consumption to meet the needs of the time
Energy Star Program
The 1995-10 International Energy Star Plan is a program developed between the United States and Japan with the goal of universally advancing energy efficient office products, such as copy machines, printers, and fax machines.
5
Phase difference between voltage and current/Correlation function with other parameters
Fluctuation state analyzed in clear waveforms.
Application example of monitor, analog, and D/A output
The 3331 and 3332 are capable of simultaneous output of voltage and current waveforms as well as the active power level, and when connected to a HIOKI MEMORY HiCORDER or Hybrid Recorder, events ranging from long-term fluctuations to transient phenomena (only with MEMORY HiCORDERs) can be recorded. Select one other measurement item (apparent power/reactive power/power factor/integration power capacity/frequency, etc.) to output from the D/A to conveniently record long-term fluctuations. 3-phase voltage/current monitor output (for the 3331)
Analog output response of the 3331/3332
Input voltage
RMS value response
0.2 seconds
MEMORY HiCORDERs To record monitor / analog / D/A output
Recording is made easy with the trigger function in the MEMORY HiCORDER, which records the rush current when a device is started, and with the recording mode, records fluctuations of power/integration values/frequency. Examination and analysis of correlation functions for each factor, including temperature, is another strong feature.
8807/8808
A compact waveform recorder the size of a B5 piece of paper for recording in 2 or 4 channels
8835-01
Waveform recorder for a maximum of 8 channels
8841/8842
Recording to a A4-width printer up to a maximum of 16 channels (8 channels with the 8841)
8826
Recording to a B5-width printer for a maximum of 32 channels
Please refer to the separate MEMORY HiCORDER catalogs
6
Basic specification
Measurement line : 3332: single-phase, 2-wire 3331: single-phase, 3-wire; 3-phase, 3-wire; singlephase, 2-wire (channel ranges cannot be independently set) Measurement item : Voltage, current, current peak, active power, apparent power, reactive power, power factor, phase angle, frequency, power integration, current integration Display indication range : 0.1% to 130% of range (zero-suppressed for less than 0.1%) (3332; zero-suppressed for less than 0.2% of V range and 40A) 3332: Voltage, current, power; Effective input range : 5% to 120% of measurement range (5% to 100% of 600V range only) 3331: Voltage, current, power; Effective input range : 1% to 120% of measurement range (1% to 100% of 600V range only) : Digital display LED, displays 4 items Display Display resolution : 99999 counts (other than integration), 999999 counts (integrated value) Rectification method : Switchable between RMS (true root mean square value) and MEAN (average rectified RMS indication). With voltage only, cutoff frequency is 500 Hz. Display update rate : Approx 5 times/sec Analog response time : FAST (0.2 sec) or SLOW (1.6 sec) (3332; at SLOW 5sec) Voltage 2M10% Current Less than 2m Voltage 600 Vrms, 1100 V peak Current 60 Arms, 90A peak 600 Vrms, 50/60 Hz Voltage (measurement range 6) / Measured value or 1100 V / measured value, whichever is lower Current (measurement range 6) / Measured value or 90 A / measured value, whichever is lower Analog output : Simultaneous output of voltage, current, active power DC5V f.s. Monitor output : Simultaneous output of voltage and current 1 Vrms f.s. : PT/CT/SC ratio Set range 0.001 to 9999 (3331; PT/CT only) Scaling : Moving average of sampling data is taken for Averaging display (1 (off), 8, 16, 32, 64 times) (3332; 1 to 300 times) Comparator (3332 only) : 2 ch (with ON/OFF function) Setting items : One item from among voltage/current/active, apparent, reactive power/power factor/phase angle/ frequency/waveform peak/integration value selected for one channel, Hi and Lo level set. : Decision and relay output (30V/0.5A) in Hi/In/Lo LED lamps. Decisions Relay Hold is possible from external control. Input resistance (50/60 Hz) Max. input voltage Max. input current Max. rated voltage to earth Crest factor : : : : :
(Time to enter accuracy range upon sudden change from 0 to 90% or 100 to 10%)
[Voltage/current/power measurement]
Measurement range : By 1-page range table
[Integration measurement]
Number of measurements : 5 times/sec Measurement range : 0.00000 to 999999 MAh/MWh (integration time up to 10,000 hours)
[Power factor/phase angle measurement]
Measurement range : -1.0000 (lead) to 0.0000 to 1.0000 (lag) -180(lead) to 0.00 to 180.00 (lag)
[Frequency measurement]
Number of channels : 1 ch Effective input range : 3332; 1 Hz to 100 kHz, 3331; 4 Hz to 50 kHz Measurement range : Auto, 500 Hz, 100 kHz (3331; up to 50 kHz)
[Wave peak measurement]
Measurement items : Displays maximum absolute current value
[D/A output]
Number of channels : 1 ch (15 bit D/A converter, polarity + 11 bits) Output resistance : 1005% Output content : Voltage, current, active / apparent / reactive power, power factor, phase angle, wave peak, frequency and the integrated value for each channel or sum of the values Output voltage : DC5V/f.s. Output update rate : 5 times/sec
[Interfaces]
GP-IB RS-232C : Conforms to IEEE-488.1 1987, with reference to IEEE-488.2 1987 : Start-stop synchronous, with baud rate of 1200 to 9600 bits/sec : External control, Display hold function, maximum value hold, current peak hold, data backup function, key lock function
[Other functions]
Measurement accuracy (at 3331: 23C3C (73F5.4F) 3332: 23C5C (73F9F)
: V, A, W Apparent / reactive : power : Integration : Power factor : Phase angle : Frequency : Wave peak
, max 80% rh, with warm-up time of at least 30 minutes, sine wave input, power factor = 1, and in-phase voltage 0) Thermal coefficient Effect of max. rated voltage to earth in-phase voltage Effect of power factor
(Period for which accuracy is guaranteed: 6 months)
Per accuracy table on page 2 1 dgt. with respect to calculation from measured value (U,I,P) sum value is max. 3dgt. 1 dgt. with respect to calculation from measured value (I,P) 1 dgt. with respect to calculation from measured value (U,I,P) 1 dgt. with respect to calculation from measured value (U,I,P) 0.1% rdg.1dgt. Measurement accuracy 1% f.s.(current peak range) Current peak range: Current range 6
: 3332; Less than 0.02%f.s./C, 3331; Less than 0.04%f.s./C : Less than 0.05%f.s. (AC 600 V rms, 50/60 Hz applied between
all input terminals and ground)
: Less than 0.4%rdg. (at 45 to 66 Hz, power factor = 0.5) Less than 0.23%f.s. (at 45 to 66 Hz, power factor = 0) Effect of external magnetic field : 1.5%f.s. (at AC 400 A rms/m, in 50/60 Hz magnetic field) : 100ppm1 sec (at 0 to 40C (32F to 104F)) Real time : Measurement accuracy0.2% f.s. D/A output Analog output : Measurement accuracy0.2% f.s. (below 45 Hz with SLOW setting) Monitor output : Measurement accuracy0.1% f.s.
General specifications
Location for use : Ambient use humidity : Ambient storage humidity : Insulation resistance : Indoors, altitude to 2000 m 0C to 40C (32F to 104F), max 80% rh (no condensation) -10C to 50C (14F to 122F), max 80% rh (no condensation) 100M or greater at DC 500 V
Dimensional drawing
3332
(8.3") (0.3") (10.3") (1.4")
Between voltage/current terminals and case, output terminals and external control terminals, voltage / current terminals and power supply, voltage terminals and current terminals, individual channels, and power supply and case Withstand voltage : AC 3.32 kV between voltage/current terminals and case, output term(50/60 Hz, 1 minute) inals and external control terminals, and between individual channels
(0.6")
Certifications
: Safety EN61010-1:2001 (Voltage and current input) Pollution factor, 2,
(3.9")
EMC EN61326:1997+A1:1998+A2:2001 Class A Power supply : AC100V to 240V 50/60 Hz (universal power supply) Maximum rated power : 3332; 40VA max. 3331; 50VA max. Dimensions and mass : 3332; Approx 210 W 100 H 261 D mm, 2.7 kg 3331; Approx 210 W 100 H 261 D mm, 2.5 kg Accessories : Power cord 1, Ext I/O male connector 1
(Approx 8.3"(W ) 3.9" (H) 10.3" (D), 95.3 oz.) (Approx 8.3"(W ) 3.9" (H) 10.3" (D), 88.2 oz.) (Not including projections such as terminals, feet, and handles)
overvoltage category III, Anticipated transient overvoltage 6000V (Power supply) Pollution factor 2 Overvoltage category II, Anticipated transient overvoltage 2500V
mass: Approx 2.7 kg (95.3 oz.)
3331
(8.3")
(0.3") (10.3")
(1.2")
(3.9") (0.6")
mass: Approx 2.5 kg (88.2 oz.)
Orders also accepted for units equipped for rack mounting. Please inquire for details.
7
Related products
Basic specification
Measurement : Single-phase, 2-wire; single-phase, 3-wire, 3-phase, 3-wire; line 3-phase, 4-wire Measurement : [When used with 9600, 9601, 9602 (option)] item voltage, current, voltage and current waveform peak, active/ reactive/apparent power, power factor, phase angle, frequency, current and power integration; load factor; efficiency [When 9603 is used (option/above functions plus functions written below)] voltage, torque, rotation count, frequency, motor output [When 9605 is used (option/all functions above plus functions written below)] harmonics, waveform, voltage fluctuation/flicker Measurement : voltage: 6/15/30/60/150/300/600/1000V range current: 200/500 mA/1/2/5/10/20/50A (when used power: 1.2 W to 150 kW (with measurement mode and voltage,
Broad bandwidth from DC/0.5 Hz to 1 MHz
Power measurement tool that supports integrated evaluations of electrical devices
In addition to basic measurement, it comes with more advanced measurement of motor power/efficiency/ harmonics /flicker (Some functions require separate options) Three types of input units can be chosen for application Simultaneous measurement of up to six lines Three types of input units can be chosen for application
frequency: 50/500/5 k/50 k/2 MHz : 0.1%rdg.0.1%f.s. (45 to 66 Hz when 9600 is used) : 8 times/s : 9600: broadband unit from DC/0.5 Hz to 1 MHz 9601: AC-only unit from 5 Hz to 100 kHz 9602: Clamp input-only unit from DC/0.5 Hz to 200 kHz Functions : Waveform peak measurement, efficiency measurement, D/A output, FDD, external control, scaling, average, backup function, motor output (Pm) measurement (9603 option), miscellaneous Power supply : AC100/120/200/230V automatic switching, 50/60 Hz, 150 VA max. Dimensions and : 430 W 150 H 370 D mm; 13 kg Mass (Approx 17.0"(W ) 6.0" (H) 14.6" (D), 459 oz.) W basic accuracy Display update Frequency quality (by unit)
with 9600) current, range combination)
3193 POWER HiTESTER
Measurement cannot be done with the 3193 unit alone. Optional input units are necessary for measurement. For more details, please request the 3193 product catalog.
3331 POWER HiTESTER (Single-phase, 3-wire and three-phase, 3-wire type) 3332 POWER HiTESTER (Single-phase, 2-wire type)
Option
9442 PRINTER
: Thermal serial dot matrix : 112 mm (4.5") : 52.5cps : 9443 AC ADAPTER or supplied nickel-hydride battery (capable of printing about 3000 lines on full charge from 9443) Dimensions and mass : Approx. 160W 66.5H 170D mm; 580 g Printing method Paper width Printing speed Power supply
(Approx. 6.3"(W ) 2.7" (H) 6.7" (D), 20.5 oz.)
9442 9443-01 9443-02 9443-03 9444 1196 9151-02 9151-04
PRINTER AC ADAPTER (For printer, Japan) AC ADAPTER (For printer, EU) AC ADAPTER (For printer, America) CONNECTOR CABLE (For printer) RECORDING PAPER (For printer, 10rolls) GP-IB CONNECTION CABLE (2m (79")) GP-IB CONNECTION CABLE (4m (158"))
Please request a 9444 CONNECTOR CABLE for connecting to the 3331/3332 unit and the 9443 AC ADAPTER when purchasing the 9442 PRINTER.
9444 CONNECTOR CABLE
Cord length approx. 1.5m
(60")
9443 AC ADAPTER
9443-02 (For the EU)
9443-01 (For Japan)
DISTRIBUTED BY
HEAD OFFICE : 81 Koizumi, Ueda, Nagano, 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail: os-com@hioki.co.jp HIOKI USA CORPORATION : 6 Corporate Drive, Cranbury, NJ 08512 USA TEL +1-609-409-9109 / FAX +1-609-409-9108 E-mail: hioki@hiokiusa.com
Shanghai Representative Office : 1310 Shanghai Times Square Office 93 Huaihai Zhong Road Shanghai, 200021, P.R.China TEL +86-21-6391-0090, 0092 FAX +86-21-6391-0360 E-mail: info@hioki.cn
All information correct as of Apr. 7, 2006. All specifications are subject to change without notice.
3330SE4-64B-03K
Printed in Japan


▲Up To Search▲   

 
Price & Availability of 3332

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X